SECTION 2: UPGRADING AN EXISTING PREP WITH AN HHS
9
10
D
Install the 7-site deck:
Plug in the Prep, power it on, and confirm that the HHS is connected:
A. R emove the deck and base plates
from their packaging.
A. Sign in, if prompted.
B. P lace the deck frame on the Prep. The sites numbered 4 and 8 should be at the front. C. Install the base plates for each site. Press down on each plate to ensure it is secure.
B. Tap New Protocol.
C. Select either “Detect” or the “Place Labware” page. D. C onfirm that site 1, the back-left site, is grayed out and says “HHS.” If the HHS does not appear, contact Hamilton support.
4 8