Hamilton Heater Shaker 3G on Prep Welcome Packet

Let’s get shaking Hamilton Heater Shaker 3G installation instructions for PREP

Hamilton Heater Shaker 3G Installation Instructions

Start your Hamilton Heater Shaker (HHS) installation in the appropriate section below.

Section 1: Installing an HHS on a New Prep.............................. 4

Section 2: Upgrading an Existing Prep with an HHS............... 10

SECTION 1:

INSTALLING AN HHS ON A NEW PREP

1

Plug in supply cable into the HHS. 2

Unpack and begin installing the Prep, following the instructions in the Prep Welcome Packet (A). The instructions will indicate to install the HHS at Step 20 ; return to these instructions at that point.

SECTION 1: INSTALLING AN HHS ON A NEW PREP

Attach mounting bracket onto Heater Shaker with four screws. 3

Place the HHS in the back-left corner of the Prep’s deck area. 4

SECTION 1: INSTALLING AN HHS ON A NEW PREP

Attach the HHS to the Prep using a 3 mm hex wrench and one of the included screws. The Prep comes with a set of hex wrenches for installation and maintenance. 5

Secure the power/ communication cable using the cable bracket and included screw. 6

SECTION 1: INSTALLING AN HHS ON A NEW PREP

Unscrew and remove the Prep’s accessory port and screw in the HHS cable. Ensure the cable is firmly attached. 7

The HHS is now installed. Continue Prep installation at Step 22 in the Prep Welcome Packet (A). 8

SECTION 2: UPGRADING AN EXISTING PREP WITH AN HHS

Power off the Prep and disconnect it from power. Installing an HHS to a powered Prep may result in damage to both instruments. 1

Remove the 8-site deck from the Prep and store it in a safe location. 2

SECTION 2: UPGRADING AN EXISTING PREP WITH AN HHS

Plug the supply cable into the HHS. 3

Attach mounting bracket onto Heater Shaker with four screws. 4

SECTION 2: UPGRADING AN EXISTING PREP WITH AN HHS

Place the HHS in the back-left corner of the Prep’s deck area. 5

Attach the HHS to the Prep using a 3 mm hex wrench and one of the included screws. The Prep comes with a set of hex wrenches for installation and maintenance. 6

SECTION 2: UPGRADING AN EXISTING PREP WITH AN HHS

Secure the power/ communication cable using the cable bracket and included screw. 7

Unscrew and remove the Prep’s accessory port and screw in the HHS cable. Ensure the cable is firmly attached. 8

SECTION 2: UPGRADING AN EXISTING PREP WITH AN HHS

9

10

D

Install the 7-site deck:

Plug in the Prep, power it on, and confirm that the HHS is connected:

A. R emove the deck and base plates

from their packaging.

A. Sign in, if prompted.

B. P lace the deck frame on the Prep. The sites numbered 4 and 8 should be at the front. C. Install the base plates for each site. Press down on each plate to ensure it is secure.

B. Tap New Protocol.

C. Select either “Detect” or the “Place Labware” page. D. C onfirm that site 1, the back-left site, is grayed out and says “HHS.” If the HHS does not appear, contact Hamilton support.

4 8

Hamilton is here for you! Visit HamiltonCompany.com/PrepHelp to browse our online help center or file a ticket.

©2023 Hamilton Company. All rights reserved. All other trademarks are owned and/or registered by Hamilton Company in the U.S. and/or other countries. Lit No: BR-2309-11 v.2.0 — 12/2023

United States +1-775-858-3000 United Kingdom, Ireland +44 (0) 121 272 92 80 Brazil +55 (11) 126 50562 China +86 21 6164 6567 France +33 184 008 420 Italy +39 39 689 33 93

Japan +81 3 6435 6850 Spain, Portugal +34 930 186 262 Denmark, Norway, Sweden, Finland +46 (0) 8 410 273 73 Germany, Switzerland, Austria, Benelux +49 (089) 248 804 808

To find a subsidiary or distributor in your area, please visit, www.hamiltoncompany.com/contacts.

Web: www.hamiltoncompany.com/robotics Email: marketingrequest@hamiltoncompany.com

Page 1 Page 2-3 Page 4-5 Page 6-7 Page 8-9 Page 10-11 Page 12-13 Page 14-15 Page 16-17 Page 18-19 Page 20-21 Page 22

Powered by