Specifications
Physical Dimensions Parameter Width without Multi-Probe Head with Multi-Probe Head
Processing Times Configuration
Microlab STARlet Microlab STAR
Task
Time(s)
Filling of one 96-well microplate with 100 µl sample (new tip for each sample)
320
1124 mm 1387 mm
1664 mm 1927 mm
8 Pipetting Channels
Aliquot reagent to a 96-well microplate (<90 µl per well)
60
Height
903 mm
903 mm
Replication of one 96-well microplate: 100 µl, new tips, with cLLD on aspiration Reformatting of four 96 well microplates to one 384-well microplate: 50 µl, new tips, with cLLD on aspiration Replication of one 384-well micro- plate: 20 µl, new tips, with cLLD on aspiration Reformatting of 4 - 384 well microplates to one 1536-well microplate: 10 µl, new tips, with cLLD on aspiration
35
Depth Manual Load Autoload
CO-RE 96 Probe Head
STARlet
781 mm 1011 mm
781 mm 1011 mm
140
Weight (8 Pipetting Channels) without Multi-Probe Head with Multi-Probe Head
135 kg 150 kg
145 kg 160 kg
35
CO-RE 384 Probe Head
< 65 dBA < 46 dBA in standby mode
Noise Level
140
(according to EN 27779)
Deck Parameters Parameter
Depth: 30.75 in 781 mm
44.25 in 1124 mm
Microlab STARlet
Microlab STAR
Deck Capacity
30 Tracks (T)
54 Tracks (T)
Modal Precision
x-y-z positional accuracy of 0.1 mm
STAR
From 10 µL to 5000 µL For detailed information on the tip size gradiations, contact your local Hamilton representative
Tip Sizes
Ambient Conditions Parameter
Operation
Storage
Transportation
Temperature
+15º C - + 35º C
- 25º C - + 70º C
- 25º C - + 70º C
Relative Humidity
15%- 85% (non-condensing)
10%- 90% (non-condensing)
10%- 90% (non-condensing)
Overvoltage Category
II
II
II
Depth: 30.75 in 781 mm
65.5 in 1664 mm
Pollution Degree
2
2
2
Altitude
up to 2000 m/6561 ft. above sea level
Indoor use only
Housing Extension Multi-Probe Head
Slim Housing Extension left
Slim Housing Extension right
Electrical Data The maximum power consumption depends on the configuration of the instrument Maximum Power Consumption 600 VA Operating Data: Voltage:
Labware All SBS standard plate types up to 1536 wells and most commercially available tube types. Contact your local Hamilton representative for more information.
100/115 VAC / 230 VAC (± 10%)
Maximum height of labware on deck: 140 mm from deck surface
Frequency:
50/60 Hz
Delayed Action Fuse:
100/115 VAC: 6.3 A (T6.3AL250) 230 VAC: 3.15 A (T3.15AL250)
Carriers For all standard labware formats and according to customer requirements. Contact your local Hamilton representative for more information.
Maximum Power Consumption 1000 VA Operating Data: Voltage:
Depth: 30.75 in 781 mm
115 VAC / 230 VAC (± 10%)
Frequency:
50/60 Hz
Depth: 30.75 in 781 mm
Depth: 30.75 in 781 mm
Delayed Action Fuse:
115 VAC: 10 A (T10AL250) 230 VAC: 5 A (T5AL250)
10.35 in 263 mm
1.15 in 29 mm
1.15 in 29 mm
36
37
Powered by FlippingBook