SECTION 1: REQUIREMENTS
First, decide on a space in your lab to put the Prep. Ensure your space fits the following specifications:
The space must have a level surface with enough room for the Prep:
Width with Touchscreen: 22.5 in / 572 mm Width without Touchscreen: 20.9 in / 531 mm Depth with Touchscreen: 25.0 in / 636 mm Depth without Touchscreen: 24.3 in / 617 mm Height: 23.8 in / 604 mm Height: (Door Open) 31.9 in / 809 mm
Allocate additional space as needed so all parts of the Prep are easily accessible.
The area should be consistently well-lit, without shadows from windows (or other instruments) that would affect labware recognition.
The area must have an ambient temperature of 15–35°C (59–95°F) with a relative humidity of 15–85%, non-condensing.
The surface must be stable and able to support the Prep’s weight (91.5 lbs / 41.6 kg). Shaking and vibrations can negatively affect your results, especially for low-volume applications.
The space must have access to electrical power between 100 VAC to 240 VAC at 50 to 60 Hz.
Once you decide on a space, you can unbox and begin installing the Prep by following the instructions in this booklet.